Thermal Calibration Equipment

  • Ellab provides both calibration equipment such as baths, dry blocks and temperature standards along with calibration services from our own metrology labs. The Ellab metrology lab facilities are among the finest in the industry since Ellab believes the starting point for any measurement is to have a reproducible and highly precise standard.

Share this product

Description

Thermal Calibration Equipment

Brand: Ellab – Denmark

Eval_calibration

Ellab provides both calibration equipment such as baths, dry blocks and temperature standards along with calibration services from our own metrology labs. The Ellab metrology lab facilities are among the finest in the industry since Ellab believes the starting point for any measurement is to have a reproducible and highly precise standard.

Ellab Temperature Standard
The traceable ETS allows for an extremely accurate temperature calibration. The ETS is available in 3 different versions and comes with calibration certificates traceable to NPL. The ValSuite™ Pro software with the ETS and the LiquiCal™ bath provide an automatic setup, which simplifies the calibration process and saves time.

 

ValSuite™ Pro Software
ValSuite™ Pro Software allows for a semi or full-automatic pre and post calibration of cable probes or wireless loggers. Using the intuitive menus, all data necessary for software calibration is processed. Offset values are generated and stored in the software. For TrackSense® Pro wireless loggers and E-Val Flex intelligent cable probes a firmware calibration routine is recommended on a regular basis, and user alerts can be programmed into the memory for smart handling.


Calibration Set-Up
Ellab’s calibration software uses a point-to-point method and can handle multiple points using only interpolation. When choosing the semi-automatic calibration, only an ETS is connected. In a full-automatic calibration, both the ETS and LiquiCal™ bath are in operation. The following performance criteria can be set:

· Maximum allowed calibration bath fluctuation band and time
· Holding time
· Max. allowed temperature standard fluctuation
· Maximum sensor temperature deviation and pass time

All setup conditions and acceptance criteria, together with actual values and deviations, are shown in the calibration report.

BATHS

LiquiCal™

LiquiCal™ is a range of high performance liquid baths used for high precision temperature calibrations. The baths can be controlled via PC allowing for automated calibration and ramp control.  It can be ran as a stand alone unit as well. LiquiCal™ provides high accuracy along with a large immersion depth, high stability, fast temperature changes and quiet runs.

LiquiCal-LL-1-web
The models cover temperatures from -80°C to +300°C, all having the same exterior dimensions, bath openings and liquid filling volume except for the micro baths (LiquiCal™ SM/HM) which are far smaller facilitating mobility.

Complete Automated Solution
The LiquiCal™ baths can be integrated with the Ellab Temperature Standard (ETS) and Valsuite™ Pro calibration and validation software for complete automated calibration and control. The software will control the ramp, stability time and automatically calibrate Ellab’s data logging equipment.

A convenient overflow reservoir captures any excess fluid that results from fluid expansion. The bath’s 15.9 L (4.2 gallon) tank can easily be emptied by drain when needed.

Liquids for LiquiCal™ Baths
To obtain optimal temperature stability and homogeneity in the calibration zone, liquids with the lowest possible viscosity (< 25 CST) are preferred. Ellab supplies a range of liquids in 3.8 L (1 gallon) containers covering the temperature range of -90°C to +300°C safely.

 
Type  Temperature
Ellab Halocarbon Oil Low -90°C to +90°C
Ellab Low Temperature Liquid -80°C to +5°C
Ellab Standard Oil -30°C to +160°C
Ellab High Temperature Oil +80°C to +300°C


One Bath/One Oil Concept
Using Ellab standard oil within a temperature range of -30°C to +150°C together with the LiquiCal™ SL bath, you will be able to cover most validation applications, including:

  • Sterilization using autoclaves and retorts
  • Pasteurization using plate heat exchangers
  • Testing using environmental chambers
  • Chilling using refrigerators
  • Storing in freezers

Ellab Temperature Standard
The traceable ETS provides an extremely accurate temperature calibration. The ETS is available in 3 different versions and comes with calibration certificates traceable to NPL. The ValSuite™ Pro software, along with the ETS and the LiquiCal™ bath, provide an automatic setup which simplifies the calibration process and saves time.

ValSuite™ Pro Software
The ValSuite™ Pro Software allows for a semi or full-automatic pre and post calibration of cable probes or wireless loggers. Using the intuitive menus all data necessary for software calibration is processed, and offset values are generated that are stored in the software. For the TrackSense® Pro wireless loggers and E-Val Flex intelligent cable probes, a firmware calibration routine is recommended on a regular basis, and a user alert can be programmed into the memory for smart handling.


Calibration Set-Up
Ellab calibration software uses a point-to-point method and can handle multiple points using only interpolation. When choosing the semi-automatic calibration only an ETS is connected. In a full-automatic calibration both the ETS and LiquiCal™ bath are in operation. The following performance criteria can be set:

  • Maximum allowed calibration bath fluctuation band and time
  • Holding time
  • Maximum allowed temperature standard fluctuation
  • Maximum sensor temperature deviation and pass time

All setup conditions and acceptance criteria together with actual values and deviations are shown in the calibration report.

LIQUICAL™ STANDARD BATHS

luidbaths

LiquiCal™ standard baths cover temperatures from -80°C to +300°C all having the same exterior dimensions, bath openings and liquid filling volume.

LiquiCal™ standard baths cover temperatures from -80°C to +300°C all having the same exterior dimensions, bath openings and liquid filling volume.

Technical Specifications

LiquiCal-LL-1-web

 
          LiquiCal LL                LiquiCal SL           LiquiCal HL
Range -80 to +110°C -45 to+150°C +35 to +300°C
Environmental +5 to +45°C / 0-90% RH
Stability at +25°C +/- 0.005°C +/- 0.005°C +/- 0.007°C
Resolution 0.01°C
Access Opening 119 x 172 mm
Immersion Depth 457 mm
Tank Capacity 15.9 L
Cooling Time
(25°C to max. cold)
Max. 3.5 Hours Max. 2 Hours
Heating Power 700 W 700 W 700 W/900 W booster
Communication RS232R
Power 230 VAC 50Hz /
230 VAC 60Hz
115 VAC 60 Hz/
230 VAX 50 Hz
115 VAC 60 Hz/
230 VAC 50Hz
Size (DxWxH):      
Internal(mm) 119x172x457
External(mm) 788x356x1067
Material (wetted parts) AISI 304
Weight 91 kg 68 kg 41 kg
Refrigerant R-507/R-508b R-507

LIQUICAL™ MICRO BATHS

LiquiCal™ micro models cover temperatures from -20°C to +200°C. Their small size facilitates mobility.

mini-bath_m_logo

DRY BLOCK

 

Dry Block Solution

When a calibration at temperatures between -100°C to +350°C (optional +700°C) is required, or when mobility is needed, a series of transportable Ellab dry block systems is available. A stable temperature profile is created using a controlled and heatedaluminum block fitted well to the sensors/probes being used. Main features are:
 

  • Easy-to-read color VGA display with perfect overview of the actual calibration status
  • Intelligent recalibration information
  • Intuitive, fast and user-friendly navigation
  • Lightweight and easy to carry around
  • Functional carrying case (included or optional)
  • Multi-hole insert kits covering all the most used sensor sizes
  • High profile design and the well-known long lasting quality

All units are featuring portability, accuracy, speed and advanced documenting functions with calibration software.

  • Temperature range: -100° to +350°C (optional +700°C)
  • Unique active dual-zone block ensures perfect temperature homogeneity in the insert, even when the calibration of large sensors or more sensors at the same time.
  • USB communication 
  • High speed heating and cooling times.

Technical Specifications

 
  PTC 155A PTC 350A RTC 156A  RTC 159A
Range

-25 to +155°C

 +33 to +350°C  -30 to +155°C  -100 to +155°C
Environmental:

0 to +40°C / 0-90% RH

0 to +40°C / 0-90% RH 0 to +40°C / 0-90% RH 0 to +40°C / 0-90% RH
Stability:

±0.01°C

 ±0.02°C  ±0.005°C  ±0.03°C
Display resolution:

1°C/0.1°C/0.01°C

1°C/0.1°C/0.01°C

1°C/0.1°C/0.01°C/0.001°C

1°C/0.1°C/0.01°C/0.001°C

Insert dimension:

25.8 x 150 mm

 25.8 x 150 mm  29.7 x 150 mm  29.7 x 150 mm
Insert capacity:

18-24 sensors +ETS

 18-24 sensors +ETS 18-24 sensors +ETS 18-24 sensors +ETS
Heating time (min to max):

16 min

  7min  19 min   26 min
Cooling time (max to min):

25 min

  24min   37 min 175 min*
Communication: USB 2.0  USB 2.0  USB 2.0  USB 2.0
Power

115/230 V 50-60 Hz

 115/230 V 50-60 Hz  115/230 V 50-60 Hz  115/230 V 50-60 Hz
External Dimensions (WxHxD):

362 x 171 x 363 mm

 362 x 171 x 363 mm  362 x 171 x 363 mm  531 x 169 x 432 mm
Weight:

10.3 kg

 8.2 kg  10.5 kg 15.2 kg
* 65 minutes from ambient to -80°C

ELLAB TEMPERATURE STANDARD

Ellab Temperature Standard (ETS)

_ETS

The traceable ETS allows for extremely accurate temperature measurements by converting resistance into temperature using its built-in algorithms.

The ETS is available in 3 different versions covering the temperature range of -196 to +420°C and comes with a calibration certificate that is traceable to NPL.

When connected to the ValSuite™ software, the LiquiCal™ bath and ETS provide a perfect setup which is automatic. This saves time and allows for a convenient calibration.

Type Temperature Accuracy
ETS10 -80 to +250°C ±0.015/0.025°C
ETS20 -50 to+150°C ±0.02°C
ETS25 -196 to +420°C ±0.025°C

Technical Specifications

 

ETS10

ETS20

ETS25

Temperature range °C

-80 to 250

-50 to 150

-196 to 420
Accuracy:
Includes Instrument, PRT and Drift:
   

-50 to 150°C

±0.015°C

±0.020°C

±0.025°C

Other temperatures

±0.025°C

NA

±0.025°C

Measuring Sample Rate

1/sec

Digital Filter

 0-10 sec

Resistance

 100Ω 

Overall Dimension

ø 11 x 475 mm

ø 11 x 475 mm

ø 11 x 475 mm

Interface & Power Source

 USB

Sensor Material

ANSI 316 

Dimension of Sensor

ø 6.35 x 300 mm

ø 6.35 x 300 mm

ø 6.35 x 300 mm

Minimum Immersion Depth  100mm 
Environmental
 15-35°C / 0 – 90% RH 
Calibration  Traceable to NPL 
Calibration Algoritms

ITS-90 

Recommended Calibration Interval

 12 months 

Weight w/wo Display  1 Kg / 0,2 Kg
Optional Display:  
Display

8 digits, 7 segments, green LED

Resolution

0.001°C

Power Supply For “Stand-Alone” Use

 12 volt 

 